Eccos®
The solution for storage and mechanical reprocessing of your Aesculap Power Systems
Basic requirements for mechanical reprocessing in combination with the Eccos® system:
- A suitable cleaner must be used (pH < 11 for Aluminum components)
- The temperature in the cleaning cycle must not exceed 55°C
- Clean for at least 3 minutes
- Apply intermediate rinse for at least 1 minute
- Carry out intensive final rinse with distilled, demineralized or fully desalinated water
- Only Thermal disinfection: 5 minutes rinse at 93°C with distilled, demineralized or fully desalinated water
- Complete the program with a drying phase of at least 10 Minutes at a temperature of not more than 120°C